diff --git a/asciidoc/volume2/gateways/tf2-ch-b-gateway-pid-mapping.adoc b/asciidoc/volume2/gateways/tf2-ch-b-gateway-pid-mapping.adoc index a5ed39d..4c65ca0 100644 --- a/asciidoc/volume2/gateways/tf2-ch-b-gateway-pid-mapping.adoc +++ b/asciidoc/volume2/gateways/tf2-ch-b-gateway-pid-mapping.adoc @@ -239,9 +239,7 @@ NOTE: <> defines the mapping of the SDC patient's sex info .R8107 [sdpi_requirement#r8107,sdpi_req_level=shall,sdpi_max_occurrence=1] **** -If <> is met and the <> is configured to export the patient's sex in the PID-8 field, then a <> / <> shall set the PID-8 field to the code for the administrative sex. - -The organization has to consider the risk that the patient's sex set in the <> and mapped to the PID-8 field does not lead to a misinterpretation of the sex concept on <> consumer side. +If <> is met and the <> provides the Healthcare Delivery Organization (HDO) the possibility to configure the export of the patient's sex set in the <> in the PID-8 field, then a <> / <> shall set the PID-8 field to the code for the administrative sex. .Notes [%collapsible] @@ -250,6 +248,12 @@ NOTE: <> defines the mapping of the SDC patient's sex info ==== **** +.R8122 +[sdpi_requirement#r8122,sdpi_req_level=shall,sdpi_max_occurrence=1] +**** +If the <> provides the Healthcare Delivery Organization (HDO) the possibility to configure the export of the patient's sex in the PID-8 field, the manufacturer of the <> shall require in the ACCOMPANYING INFORMATION that the HDO has to consider the risk that the patient's sex set in the <> and mapped to the PID-8 field does not lead to a misinterpretation of the sex concept on <> consumer side. +**** + [#ref_tbl_dec_obx_sex_mapping] .OBX Sex Mapping |===