Apply hysteresis and tau correction to Oxygen SBE43 data #5

Open
ghost opened this issue Apr 2, 2019 · 3 comments
Open

Apply hysteresis and tau correction to Oxygen SBE43 data #5

ghost opened this issue Apr 2, 2019 · 3 comments

Comments

@ghost
Copy link

ghost commented Apr 2, 2019

No description provided.

@ghost ghost self-assigned this Apr 2, 2019
@ghost ghost added the SBE19plusV2 label Apr 2, 2019
@ghost
Copy link
Author

ghost commented Apr 2, 2019

Currently for the tau correction I am only calculating dvdt between the prior and current voltage values. Seabird indicates that this should be calculated using a 2 second window, which with a sampling rate of 4hz, would be 8 voltage readings.

@ghost
Copy link
Author

ghost commented Apr 2, 2019

For the hysteresis correction, the equation is listed in the Seabird Application Note 64.3, p. 6

@ghost
Copy link
Author

ghost commented Apr 10, 2019

In talking with Seabird personnel, the tau correction dv/dt value is actually calculated as:

It's the change in volts from the SBE43 over the change in time it takes the instrument to travel 2 decibars during the profile.