EecSeq

EecSeq Library Prep

The Expressed Exome Capture Sequencing protocol is designed to create exome capture probes directly from RNA. The probes are then used from hybrid capture of exome DNA sequences, allowing for genotyping of alleles at expressed genes.

WARNING: These protocols are abridged versions of the manufacturers recommended protocols. Use of these protocols assumes familiarity with manufacutrers instructions. Please read manufacturers instructions if this is your first time (or first time in a while) performing these procedures.

RNA Prep

  • Takes 1.5 hours
  • Takes 0.5 hours
  • Takes 1/2 day to get results back. Can only get bioanalyzer runs on M, T, W.
  • Takes 45 minutes
  • Anneal the SAI adapters designed for Illumina TruSeq. The Pinsky Lab uses an SAI1 adapter and an SAI1-no-N adapter. Use one or the other equally among samples.
  • Takes 6 hours
  • There are safe stopping points within this protocol.
  • Takes 1/2 hour.

Safe Stopping Point

This is a safe stopping point. If you are stopping for more than a week, store your sample at ‐15° to ‐25°C.

  • Takes 6 hours (5 hours on thermocycler)
  • Takes 1 hour

Safe Stopping Point

This is a safe stopping point. If you are stopping, store your sample at ‐15° to ‐25°C.

  • Takes 1/2 hour.
  • Use a sample volume of 80ul and an ampure volume of 128ul.
  • Conduct the cleanup in a 1.7mL tube.
  • Elute 25ul into a 200uL tube using 10 mM Tris-HCl (pH 8.0).
  • Takes 1/2 hour plus thermocycler time.
  • Takes 1/2 hour
  • Use a sample volume of 50ul and an ampure volume of 80ul.
  • Conduct the cleanup in a 1.7mL tube.
  • Elute 22 μl of 10 mM Tris-HCl (pH 8.0) into a 200uL tube.
  • Takes 1/2 hour.

Split finished cDNA library for each sample into two vials
- One tube for sequencing
- One tube for probe synthesis

Note that this step is optional as sequening the probes directly is not necessary for exome capture analysis. Alternatively, a portion of the mRNA library can be saved before DSN normalization for sequencing.

Safe Stopping Point

This is a safe stopping point. If you are stopping, store your sample at ‐15° to ‐25°C.

Probe Synthesis

  • Takes 17 hours, 16 is on thermocycler (overnight).
  • Takes 1/2 day to get results back. Can only get bioanalyzer runs on M, T, W.
  • Check against the bioanalyzer results in the Puritz protocol to make sure everything looks the way it should.

Safe Stopping Point

This is a safe stopping point. If you are stopping, store your sample at ‐15° to ‐25°C.

  • Takes 1/2 hour
  • use 45ul of sample and 81ul of ampure.
  • Conduct the cleanup in a 1.7mL tube.
  • Elute 20 μl of 22ul of 10 mM Tris-HCl (pH 8.0) into a 200uL tube.

Safe Stopping Point

This is a safe stopping point. If you are stopping, store your sample at ‐15° to ‐25°C.

  • Takes 1/2 hour
  • use 22ul of sample and 33ul of ampure.
  • Conduct the cleanup in a 1.7mL tube.
  • Elute 30 μl of 31ul of 10 mM Tris-HCl (pH 8.0) into a 200uL tube.
  • Takes 21 hours, 20 hours on heat block.
  • Takes 1/2 hour
  • use 50ul of sample and 75ul of ampure.
  • Conduct the cleanup in a 1.7mL tube.
  • Elute 20 μl of 10 mM Tris-HCl (pH 8.0) into a 200uL tube.
  • Takes 1/2 hour
  • Takes 1/2 day to get results back. Can only get bioanalyzer runs on M, T, W.
  • Check against the bioanalyzer results in the Puritz protocol to make sure everything looks the way it should.

Safe Stopping Point

This is a safe stopping point. If you are stopping, store your sample at ‐15° to ‐25°C.

Genomic Library Prep

  • Takes 1 hour.
  • There is a specific set of adapters made for the EecSeq project.
  • Takes 22 hours with a 16 hour overnight lysis step.
  • elute DNA into 35ul water or Tris-HCl and continue immediately to next step. Do not use an EDTA containing solution like TE to elute.
  • Takes 2 hours.
  • EecSeq protocol says to use 1/2 the reaction volume recommended by KAPA.
  • Takes 2 hours.
  • Takes 3 hours.
  • Takes 1/2 hour
  • use 55ul of sample and 44ul of Kapa Pure Beads.
  • Conduct the cleanup in plates.
  • Elute 11 μl of 12.5ul of 10 mM Tris-HCl (pH 8.0) into a fresh plate.
  • Takes 1/2 hour
  • Takes 1 hour
  • Takes 1/2 hour plus thermocycler time
  • Takes 1/2 hour
  • use 25ul of sample and 25ul of Kapa Pure Beads.
  • Conduct the cleanup in a 1.7mL tube.
  • Elute 15ul of 10 mM Tris-HCl (pH 8.0) or water into a screw cap tube if stopping.

Safe Stopping Point

This is a safe stopping point. If you are stopping, store your sample at ‐15° to ‐25°C.