Spine Scanning Protocol

BenChurch edited this page Oct 2, 2017 · 1 revision

Tools needed:

  • Sonix Touch ultrasound machine
  • C5 gps probe
  • One Model 800 sensors (large white brick)
  • Computer running Slicer connected to Sonix Touch through local network
  • Make sure SlicerIGT extension is installed in Slicer
  • Ultrasound gel
  • Paper towels

Workflow

  • Connect two sensors to the tracker control box: Plug probe sensor to Transducer 1, plug Reference sensor to Transducer 2.
  • Start PlusServer application on Sonix Touch. Make sure version number contains "4412"
  • Select config file in Plus Server: S:\data\Scoliosis\ConfigFiles\Spine_C5_d100_v01.xml
  • Start Slicer on the desktop computer
  • Create OpenIGTLink connection: IP = 130.15.7.75, port = 18944 (These are displayed in Plus Server.)
  • Go to PlusRemote module in Slicer
  • Click >> button next to "Start recording" to access recording options
  • Change file name according to e.g. Scan01-Top_Subject04.mha
  • Check "Add timestamp to filename"
  • Volunteer reads and signs consent form
  • Check if the volunteer already has a subject number in our shared Excel table
  • Check your operator number in the same Excel table
  • Ask volunteer to remove or open clothes on the back
  • Place reference sensor on the belt or back of waist. Make sure orientation is always the same.
  • Ask volunteer to put forehead and toes against the wall, and push the shoulders forward
  • Make sure there is enough gel on the back
  • Find the T1 level (just below the prominent C7 vertebra)
  • Scan and save data in two portions. Stop and save approximately halfway on the back to apply more gel.
  • On Sonix Touch go to D:/PlusPackages/PlusApp-2.3.0.4412.../Data
  • Move recorded data on the S: drive. One scan session, one volunteer is going to one folder. Name the folder as 2016-08-15_Op04_Subject04. Where Op04 is the operator number, and Subject04 is the volunteer number (both in shared Excel table).
Clone this wiki locally