Ion Pairing

Sadjad F Baygi edited this page Nov 1, 2022 · 7 revisions

Ion Pairing

The ion pairing approach embedded in the IDSL.IPA allows to significantly decrease the number of candidate m/z for chromatography analysis and accelerate the workflow. The noise level prior to ion pairing is only applied to the most abundant ions in individual chromatogram scans. By default, IDSL.IPA screens for carbon-related isotopologues (ΔC = 13C - 12C = 1.003354835336 Da) since carbon-containing compounds are targeted almost in the entire mass spectrometry analyses.

Noted key advantages of the 12C/13C isotopologue pairing are

  1. reducing false positive peaks
  2. obviating the postprocessing 13C deisotoping step
  3. measuring peak quality by number of scans with confirmed 12C/13C isotopologue pairs within a chromatography peak (nIsoPair)
  4. prediction of number of carbons for elemental composition using the R13C = 1.0816Cn equation, where Cn indicates number of possible carbons, Cn = 92
  5. evaluating the accuracy and specificity of a peak for an organic compound across multiple samples since we the ratio of ions for a true positive should remain steadily constant

Ion pairing with satellite ions

You still can search for negative mass differences ions as shown below. This option allows to only focus on the ion pairs of interest and exclude non-related ions. Ratio of Satellite Ion (RSI) is calculated similar to R13C and may be used and adjusted using PARAM0025.